Siemens ADVIA 1800 Operator's Manual page 148

Chemistry system
Table of Contents

Advertisement

1. Set the analyzer to Standby.
2. Open the filter cap at the front of the cuvette conditioner bottle (
tube with the filter.
3. Disconnect the cuvette conditioner bottle level sensor connector (
4. Turn the connector counter-clockwise and pull it out.
5. Remove the bottle.
6. Empty any remaining contents of the bottle.
7. Rinse the bottle with deionized water and drain well.
8. Refill the bottle with cuvette conditioner.
9. Return the bottle to the same position on the shelf in the cabinet.
10. Connect the cuvette-conditioner bottle level sensor connector, then push the
connector in and turn clockwise.
11. Insert the filter and hose, then fasten the cap.
NOTE:
12. Prime the fluid lines:
NOTE:
a. Return the analyzer to the READY mode.
b. At the Operation Panel, select
c. At the Prime Settings dialog box, select
times in all fields.
d. Select
148
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
CAUTION
Ensure that the level-sensor connector (
Make sure that the filter holder is located at the bottom of the bottle.
If you are cleaning other bottles, wait to perform this step for all fluid lines.
.
Execute
) does not get wet to avoid damaging it.
2
.
Prime
and type
Prime 2
ADVIA 1800 Operator's Guide
), and pull up the
3
).
2
for the Number of
10