Every 3 Months Maintenance; Replacing The Lamp - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

Every 3 months maintenance

Replacing the lamp

Materials required:
• Halogen lamp, 12 V/50 W
(REF 02127928, PN 073-0099-01)
Time: 60 minutes
Analyzer mode: STANDBY
The lamp MUST be replaced:
Quarterly
After approximately 2000 hours of use
If the system warns that lamp energy is out of range
If the Checking Lamp Energy procedure indicates the A-D points are outside the ±40
range of the scatter plot center line or the volts for any of the wavelengths are outside
the range of 5.0 to 9.0 volts
If the ATTENU(%) for any of the 14 wavelengths on the Lamp Check Energy
window falls below 80%
Figure 5-22. Components of the lamp
1. Set the analyzer to
2. Lift and remove the access panel in front of the Rotating Reaction Tray to expose the
lamp housing.
Maintenance
.
Standby
WARNING
The lamp housing is
hot
(approximately 10 minutes) before touching any components.
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
Lead Wire Connector
2
Lead Wire Connector
3
Lamp Screws
4
Lamp Plate
5
Alignment Hole and Pin
. To avoid burns, allow it to cool down
149