• Sample information and system status
The button bar at the top of the window displays the Sample Search and Rack or LAS.
Sample Info. buttons.
• Press the
sample information by sample number by STT position (tray number and sample
position), or by rack/LAS position (rack number and sample position).
The search returns a Sample Information window showing the sample number, position
number, sample status, and the time remaining to complete the analysis of the sample.
• If a rack handler or laboratory automation system (LAS) is in use, select the
LAS. Sample Info.
LAS. Sample Information window:
• Sample barcode number.
• Sample Status (see the status codes listed on the Test Result Monitor window).
• The time remaining to complete the processing of the sample.
If processing has completed or if there are no samples available for processing
NOTE:
when you select the Rack or LAS. Sample Info. button, the system displays a message
that no sample information is available.
The left side of the Test Result Monitor window displays three panels.
• The System Status panel shows the current operating mode of the system.
• The Sample information panel shows the sample number and sample position of the
currently selected position on the STT/CTT graphic. If this is a barcode analysis, the
barcode number displays, and the position number is 0-00.
• The code panel shows the color codes used in the STT/CTT graphic to represent
sample status.
In the middle of the sample tray display, the tray (TT) number for the current run displays
in the TT No. list box. If this is a barcode analysis, the TT number is 0.
To view the status of prior tray samples that are still in process, select the down arrow of
the TT No. list box, then select the number of the tray you want to view.

Review/Edit window

CAUTION
Tests on the Review/Edit window must be in the same order as they display at the
System Test List window to avoid incorrect positioning of results on the print
report.
after the initial setup. Any repositioning of test names on the System Test list
window after initial setup could result in associating already-run tests with the
wrong test name.
Perform any of the following tasks at this window:
• review sample results
• configure reruns
• print selected patient results
• transmit selected patient results
44
button to display a dialog box where you can search for
Sample Search
button to display the following sample information in the Rack or
Do not modify the order
of the test names at the System Test list window
ADVIA 1800 Operator's Guide
Rack or