During operation system settings area
1. In the Realtime monitor area, select from the following formats for routine analysis:
Standard
Conc.
2. In the On-line field, select
3. In the ADVIA QC Transfer area, select
4. In the syngo Transfer field, select
Pre-operation settings
NOTE:
system is initialized, the options are not available on the screen
Specify if these system components are operational or cancelled:
• Chemistry
• ISE
• Unused
• Labo.Auto.Sys./Rack Handler
Clot Detection settings
1. In the Clot detect area, select
disable it.
2. In the Clot sensitivity area, select
clot detection sensor.
Setting up the ISE
This section describes the steps required if the present configuration for the electrolyte
analyzer (ISE) is ever lost or requires modification.
Activating the ISE
1. Set the ELA switch in the power supply distribution chassis panel (on the back of the
analyzer) to the
2. At the Systems Specifications Settings window, set the Electrolyte option to
3. At the System Monitor window, set the electrolyte option to
190
Conc. (ABS)
Detail
Do
You can change these options only before you initialize the system. After the
position.
ON
Conc. (ABS – RB)
or
.
Not do
or
.
Yes
No
or
Yes
No.
. to use the clot detection feature or N.A. to
Avail
,
, or
Low
Middle
High
.
to select the sensitivity of the
Avail
.
Operate
ADVIA 1800 Operator's Guide
.