Performing additional ISE electrode washes

Materials required:
• ISE Detergent Solution
(REF 01307361, PN B01-4174-01)
Time: 5 minutes
Analyzer mode: Manual operation
ISE Detergent Solution is automatically run through the ISE module as part of the
shutdown wash procedure (WASH2). Manually perform additional ISE washes
(described in the following procedure) more frequently, but do not exceed 3 times in 24
hours in the following circumstances:
• Dialysis samples are run routinely
• Veterinary samples are run routinely
• The system is in use more than 8 hours a day in a high volume laboratory,
which is having persistent imprecision or Chloride bias issues that have not
been resolved by any other means.
Maintenance
2
2
: At your laboratory's discretion, you may use other positions for the washes
.
WASH2
for Cycles.
2
in the CTT cup position 1st time field and
49
2nd time field.
in the RTT1 and RTT2 cup positions 1st time fields and
55
and RTT2 cup positions 2nd time fields.
.
Execute
) and RTT2 (
3
) position #55 contains a 10% solution of
) position #50 contains DI water.
1
) and RTT2 (
) position # 56 contains DI water.
3
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
WARNING
Read and follow the cautions on
the box and label before
handling any reagents. Some
method and system reagents are
classified as hazardous material.
in the CTT cup position
50
in the RTT1
56
129