Performing the startup wash (WASH3)

Materials required:
• Deionized water
Time: 26 minutes
Analyzer mode: READY
The daily startup wash rinses the probe lines, reaction cuvettes and dilution cuvettes.
Location
1
1
1
1
1
2
2
2
2
3
3
3
3
Figure 5-12. Location of wash solutions on the CTT and RTT trays
Maintenance
Position
Wash Solution
CTT-15
ISE Detergent Solution
CTT-16
Deionized water
CTT-49
10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
CTT-50
Deionized water
CTT-51
Deionized water
RTT1-53
Reagent Probe Wash 1
RTT1-54
Reagent Probe Wash 2
RTT1-55
10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
RTT1-56
Deionized water
RTT2-53
Reagent Probe Wash 1
RTT2-54
Reagent Probe Wash 2
RTT2-55
10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
RTT2-56
Deionized water
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
127