Cleaning and replenishing the dilution bottle

Materials required:
• Deionized Water
• Physiological saline (0.9% NaCl)
Time: 10 minutes
Analyzer mode: STANDBY
NOTE:
once a month.
WARNING
Figure 5-16. Saline diluent bottle
1. Set the analyzer to Standby.
2. Lift the cover from the saline diluent bottle (
3. Empty the remaining contents of the bottle.
4. Rinse the bottle with deionized water and drain well.
5. Refill the bottle with 0.9% saline diluent.
6. Replace the bottle in the same position on the shelf in the cabinet.
7. Replace the cover of the diluent bottle.
NOTE:
the dilution bottle.
8. Prime the fluid lines:
NOTE:
can prime all the fluid lines at once.
Maintenance
The dilution bottle may be cleaned when it is refilled, but must be cleaned at least
Before handling any reagents, read the warnings on page 196
CAUTION
Note the bottle position on the shelf to avoid mixing up the fluid bottles.
Make sure that the Teflon tube and filter holder are located at the bottom of
If you are cleaning the detergent or cell conditioner bottles at this time, you
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
Saline Diluent Bottle
2
Level Sensor Connector
), and remove the bottle.
1
141