8 System Setup

Connecting the ADVIA Chemistry System and your host
computer
1. Have your laboratory computer professional connect the host computer as described
in the publication titled LIS Interface Guide.
2. Use the Online Settings window to enter the communication parameters.
3. Use the Online settings window to determine which sample results and workorders to
transmit.
This feature allows you to transmit results when they are available. This is called
"real-time transmission."
If you do not want to use real-time transmission, you can always transmit the results
as a batch using the Review/Edit window.
Similarly, you can download workorders at any time as a batch using the Order Entry
window.
4. Use the Item Setting dialog box at the Online Setting window to determine which test
results to transmit and to assign host test numbers.
5. Use the Data Clean Setting dialog box at the Online Settings window to have the
system automatically validate the sample results before they are transmitted.
Use the Transfer Results List dialog box of the Review/Edit to review the data clean
report for each sample.
6. In the System basic composition area at the System Specification Settings window,
select
7. In the During-operation system set area of the System Monitor window, select
the Online option.
8. Use the Host button in the Operation Panel to disconnect from or reconnect to the
host computer.

Setting the System Monitor

1. At the Menu Panel, select
2. At the System Monitor window, enter or change the settings defined below, as
needed.
System Setup
Avail
.
Maint
., then select
System Monitor
for
Do
.
189