to take the first measurement point.
4. Press
ENTER
After taking measurement point, the clock hour hand moves automatically to
the next clock position. (If the first measured position is 12:00 o'clock, then the
hour hand moves to the 1.30 o'clock position).
5. Rotate the shafts to the next measurement position (this could be either 1:30,
3.00, 4.30, 6.00, 7.30, 9.00 or 10.30 o'clock position).
If shaft rotation restrictions hinder the taking of measurement at particular
shaft positions, bypass these using the navigation keys.
6. Press
to take measurement point.
ENTER
A minimum of three measurement positions are required to obtain results, but
it is recommended that a maximum number of measurement points over a
wider rotational angle be taken to enhance the diagnosis.