Thermocouple Calibration - Omron E5C User Manual

Digital temperature controllers
Hide thumbs Also See for E5C:
Table of Contents

Advertisement

7 User Calibration
7-3

Thermocouple Calibration

• Calibrate according to the type of thermocouple: thermocouple group 1 (input types 5, 7, 11, 12, 15,
19, and 20) and thermocouple group 2 (input types 6, 8, 9, 10, 13, 14, 16, 17, 18, 21, 22, 23, and 24).
• When calibrating, do not cover the bottom of the Controller. Also, do not touch the input terminals
(terminals 5 and 6 on the E5CC, terminals 1 and 2 on the E5CC-U, terminals 23 and 24 on the
E5EC/E5AC, or terminals 13 and 14 on the E5DC) or the compensating conductors.
Preparations
11
Input power supply
12
1
Input power supply
2
• Set the cold junction compensator designed for compensation of internal thermocouples to 0°C.
Make sure that internal thermocouples are disabled (i.e., that tips are open).
• In the above figure, STV indicates a standard DC current/voltage source.
• Use the compensating conductor designed for the selected thermocouple. When thermocouples R,
S, E, B, W, or PLII or an infrared temperature sensor is used, the cold junction compensator and the
compensating conductor can be substituted with the cold junction compensator and the
compensating conductor for thermocouple K.
Additional Information
Connecting the Cold Junction Compensator
Correct process values cannot be obtained if you touch the contact ends of the compensating
conductor during calibration of a thermocouple. Accordingly, short-circuit (enable) or open
(disable) the tip of the thermocouple inside the cold junction compensator as shown in the figure
below to create a contact or non-contact state for the cold junction compensator.
E5@C
Compensating conductor
7 - 4
E5CC
5
Cold junction
compensator
+
6
0°C/32°F
Compensating conductors
E5EC/AC
23
Cold junction
compensator
+
24
0°C/32°F
Compensating conductors
Cold junction compensator
Short-circuit
0°C/32°F
STV
10
Input power supply
11
STV
11
Input power supply
12
Cold junction compensator
0°C/32°F
E5@C
Compensating conductor
E5@C Digital Temperature Controllers User's Manual (H174)
E5CC-U
STV
2
Cold junction
compensator
+
1
0°C/32°F
Compensating conductors
E5DC
STV
13
Cold junction
compensator
+
14
0°C/32°F
Compensating conductors
Open

Advertisement

Table of Contents
loading

This manual is also suitable for:

H174-e1-05 series

Table of Contents