The measured values are displayed on the screen.
3.8.2

DC or AC+DC Measurement

Set the switch to
zero must be corrected first.
Step 1: Correction of DC Zero
Important: The clamp must not be closed on the conductor during the DC zero
correction. Hold the clamp in the same position during the whole procedure so
that the correction value will be exact.
Press the
button until the device emits a double beep and displays a value
near "0". The correction value is stored until the clamp is powered down.
NOTE: The correction is effected only if the value displayed is < ±20A, otherwise
the value displayed blinks and is not stored. The clamp must be recalibrated (see
§ 5.3)
Display
1st row
RMS current A RMS
2nd row
Crest factor (CF)
3rd row
DC current component A
and select DC if the display does not indicate "0"; the DC
Quantity
DC
28