AEMC F03 User Manual

Clamp-on multimeter
Table of Contents

Advertisement

Quick Links

F03
CLAMP-ON MULTIMETER
E N G L I S H
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for AEMC F03

  • Page 1 CLAMP-ON MULTIMETER  E N G L I S H User Manual...
  • Page 2 The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com. Serial #: ________________________________ Catalog #: 2129.52 Model #:...
  • Page 3: Table Of Contents

    Table of Contents 1. INTRODUCTION ................3 1.1 International Electrical Symbols ..........3 1.2 Definition of Measurement Categories ........4 1.3 Receiving Your Shipment ............4 1.4 Ordering Information ..............4 1.4.1 Accessories and Replacement Parts ......4 2. PRODUCT FEATURES ..............5 2.1 Description ................5 2.3 Rotary Switch Functions ............7 2.4 Command Buttons ..............7 2.5 Hold Button Primary Functions ..........7 2.5.1 Display Lock ..............7 2.5.2 Preselecting Min/Max Mode ........7 2.5.3 Automatic Compensation for Lead Resistance ....8...
  • Page 4 2.11 Liquid Crystal Display ............. 11 2.11.1 Digital Display ............. 11 2.11.2 Symbol Display ............12 2.12 Buzzer ..................13 3. SPECIFICATIONS................14 3.1 Reference Conditions .............14 3.2 Electrical Specifications ..........14 3.2.1 Voltage ( ) ..............14 3.2.2 Continuity ( ) .......... 15 3.2.3 Resistance (Ω) ............15 3.2.4 Semi-Conductor Test ( ) ..........15...
  • Page 5: Introduction

    Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous. This symbol refers to a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is permitted. In conformity with WEEE 2002/96/EC Clamp-on Multimeter Model F03...
  • Page 6: Definition Of Measurement Categories

    Ordering Information Clamp-on Multimeter Model F03 ........Cat. #2129.52 Includes multimeter, set of red and black leads with probe tips, 1 K-type thermocouple adapter, 9V battery, carrying pouch and user manual. 1.4.1 Accessories and Replacement Parts Replacement set of leads, red and black with probe tips ..
  • Page 7: Product Features

    CHAPTER 2 PRODUCT FEATURES Description The Clamp-on Multimeter Model F03 emphasizes reliability and simplicity of use to respond to the needs of power professionals. Features: TRMS measurement  A compact unit, integrating the current sensor for  intensity measurements without breaking the test circuit Outstanding ergonomic features:  ● automatic selection of the type of signal to be measured, AC or DC ● automatic selection of AC or DC measurement ● automatic selection of measurement ranges ● programmable audio voltage indication (V-Live) ● “over-range” indication ● backlighting of the digital display ● power auto-off ● MIN - MAX - PEAK value recording function ● correction of differences in DC measurement (DC zero) ● automatic compensation of measurement lead resistance (Ω zero) Compliance with IEC electrical safety standards and ...
  • Page 8 Model F03 Callouts   Jaws Command Buttons  5-way Rotary Switch  Liquid Crystal Display Clamp-on Multimeter Model F03...
  • Page 9: Rotary Switch Functions

    Hold Button Primary Functions 2.5.1 DISPLAY LOCK Short-press the HOLD button to freeze/lock the display. Press again to unlock. 2.5.2 PRESELECTING MIN/MAX MODE Short-press the HOLD button, then the MIN/MAX button to prese- lect the MIN/MAX mode. Press the HOLD button again to make the MIN/MAX mode effective. (e.g., use this function to preselect the MIN/MAX mode to prevent unwanted or mistaken integration of MIN/MAX values.) Clamp-on Multimeter Model F03...
  • Page 10: Automatic Compensation For Lead Resistance

    While pressing down the HOLD button, bring the rotary switch from the OFF position to the position. The unit emits a double beep, then the V and symbol flashes. The selected configuration is put into memory when the button is released (the V symbol becomes fixed and the symbol flashes). Proceed in the same way to deactivate the V-Live function (the symbol disappears when the button is released). Clamp-on Multimeter Model F03...
  • Page 11: Displaying The Internal Software Version

    Short-press the yellow button to select °C or °F. The selection made will not be saved when the clamp is turned off. Yellow Button Secondary Functions (with rotary switch) 2.8.1 MODIFICATION OF AUDIO INDICATION THRESHOLD IN CONTINUITY TEST While pressing down the yellow button, bring the rotary switch from the OFF position to the position. The unit beeps, the Ω and symbols appear, along with the threshold value (40.0 by default). Clamp-on Multimeter Model F03...
  • Page 12: Selecting °C Or °F As The Default Setting

    While pressing down the yellow button, turn the rotary switch from the OFF position to position. The unit emits a double beep, then all the segments of the digital display and the symbol flashes. The default configuration is set when the button is released (the display no longer flashes and the symbol disappears). The default configuration is: • Audio identification threshold: 40Ω • Auto-off: ON • V-Live function: none • Unit for temperature measurement: not controlled Clamp-on Multimeter Model F03...
  • Page 13: Min/Max Button Functions

    Held pressure: Display of estimated remaining battery power, in hours. 2.11 Liquid Crystal Display The liquid crystal display includes the digital display of the measured values, the related units and symbols. 2.11.1 DIGITAL DISPLAY 4 digits, 9999 counts, 3 decimal points, + and - signs (DC and peak mea- surement) + OL : Positive value range exceedance (>3999cts) - OL : Negative value range exceedance OL : Unsigned value range exceedance - - - - : Indeterminate value (middle segments) Clamp-on Multimeter Model F03...
  • Page 14: 2.11.2 Symbol Display

    2.11.2 SYMBOL DISPLAY Flashing: power limited to approximately 1 hour Steady: battery drained, operation and accuracy are no longer guaranteed Constant operation (no automatic shutdown) Fixed: Continuity measurement Flashing: V-Live function selected HOLD Function active PEAK ON in V and A in MIN/MAX mode if the measurement of the peak value is selected MAX Indicates the display of a maximum value in MIN/MAX mode MIN Indicates the display of a minimum value in MIN/MAX mode Fixed: measurement in AC manual mode Flashing: measurement in AC automatic mode Fixed: measurement in DC manual mode Flashing: measurement in DC automatic mode T° Temperature measurement INT Measurement of temperature when the terminals are not connected, or the thermocouple connected is faulty. EXT Measurement of temperature when the thermocouple is connected Semi-conductor test on position Ω Clamp-on Multimeter Model F03...
  • Page 15: 2.12 Buzzer

    2 short, high-pitched sounds: validation of a configuration parameter • Short and medium sound every 400 ms: voltage measured is higher than the unit’s guaranteed safety voltage • 5 short and medium recurring sounds: automatic shutdown of the instrument • Continuous medium sound: measured continuity value lower than programmed threshold, short-circuit connection during semi- conductor test • Modulated medium continuous sound: value measured in volts, higher than 45V peak when the V-Live function is selected Clamp-on Multimeter Model F03...
  • Page 16: Specifications

    In AC, the display indicates OL over 600Vrms (900V in PEAK mode). **In AC if the value of the voltage measured is <0.15V the display indicates 0.00. MIN/MAX Mode: Accuracy: same as previous table +0.2% of Reading Capture Time: 100 ms typical PEAK Mode: Accuracy: same as previous table +2% of Reading Capture Time: 500 µs typical (2.5 ms max) Detection Threshold Accuracy (V-Live Mode): 45V peak ± 2V Clamp-on Multimeter Model F03...
  • Page 17: Continuity ( )

    Overload Protection 500V or 750V or peak *With compensation for measurement cable resistance MIN/MAX Mode: Accuracy: same as previous table +0.2% of Reading Capture Time: 100 ms typical 3.2.4 Semi-Conductor Test ( Display Range Measuring Range 0 to 3.199V Accuracy 1% of Reading + 2cts Resolution Measuring Current* 2mA to 4mA Overload Protection 500V or 750V or peak *Per the voltage measured Clamp-on Multimeter Model F03...
  • Page 18: Current ( )

    +400°F to +120.0°F +399.9°F +1832°F ±1.5°C 1% of R ± 1.5°C 1% of R ±1.5°C Accuracy ±2.7°F 1% of R ± 2.7°F 1% of R ±2.7°F 0.1°C 1°C Resolution 0.2°F 1°F Detection of sensor cutoff – INT symbol lit instead of EXT Thermal time constant 0.7 min/°C According to the sensor model Note: The accuracy stated for external temperature measurement does not include the accuracy of the K couple. Clamp-on Multimeter Model F03...
  • Page 19: Power Supply

    Power Supply Battery: Standard 9V alkaline (type IEC 6LF22, 6LR61 or NEDA 1604) Charge life: 75 h or 25,000 x 10 s measurements Low Battery indicator: Flashing: Charge life < 1 h Fixed: Change battery Auto-off: 10 minutes with no action on the rotary switch or the buttons Mechanical Specifications 1. Reference Range 2. Operating Range Temperature: 3. Storage Range (without battery)  Œ Ž Ž Temperature in °C Operating Temperature: 32 to 122°F (0 to 50°C); 90% RH Storage Temperature: -40 to 158°F (-40 to 70°C); 90% RH Altitude: Operation: ≤2000m Storage: ≤12,000m Dimensions: 2.76 x 7.6 x 1.46" (70 x 193 x 37mm) Weight: 9.17 oz (260g) Clamp Tightening Capacity: ≤1.00” (≤26mm) Clamp-on Multimeter Model F03...
  • Page 20: Safety Specifications

    • Electrostatic discharges: 4kV on contact, aptitude criterion B 8kV in the air, aptitude criterion B • Radiated field: 10V/m, aptitude criterion B • Fast Transients: 1kV, aptitude criterion B • Conduit interference: 3V/m, aptitude criterion A Mechanical Resistance • Free fall 1m (test as per IEC 68-2-32) • Impacts: 0.5 J (test as per IEC 68-2-27) • Vibrations: 0.75mm (test as per IEC 68-2-6) Auto Power OFF (per UL94) • Housing V0; Jaws V0; Display window V2 Clamp-on Multimeter Model F03...
  • Page 21: Variations In Operating Range

    50 dB 45 dB Rejection of serial 0 to 600V <1ct 60 dB mode in AC 0 to 400A 55 dB 40 dB Rejection of common <1ct 60 dB 0 to 600V/50Hz mode <1ct 0.1A Influence of external 0 to 400A/m 65 dB 60 dB magnetic field (50Hz) Number of moves 50,000 0.3% R 1% + 1ct opening of jaws Clamp-on Multimeter Model F03...
  • Page 22 Typical Frequency Response Curves - V = f (f) Specified limits -10% 10000 Hz 100 Hz 10 Hz 1000 Hz Frequency (Hz) - I = f (f) Specified limits 10000 Hz 10 Hz 100 Hz 1000 Hz Frequency (Hz) Clamp-on Multimeter Model F03...
  • Page 23: Operation

    • Range switching and AC/DC selection are automatic. • Short-press the yellow button to manually select AC/DC If the signal measured is >45V peak, the audio indication is acti- vated if the V-Live function is selected (see § 2.6.2). For voltages ≥600V or Trms, a repetitive beep of the buzzer indicates that the measured voltage is higher than the acceptable safety voltage (OL). Audio Continuity Test - 1. Connect the measurement cables to the terminals. 2. Set the rotary switch to the “ ” position. 3. Connect the unit to the circuit to be tested. The buzzer is contin- uously active as soon as contact is established (circuit closed) and if the resistance value measured is lower than the threshold value chosen by the programming (adjustable from 1 to 40Ω,). NOTE: Above 400Ω, the display indicates OL. Clamp-on Multimeter Model F03...
  • Page 24: L Ead Resistance Compensation (Ω Zero)

    “+” terminal and black lead on the “COM” terminal. 2. Set the rotary switch to the “ ” position and press twice on the yellow button: The symbol is displayed. 3. Connect the unit to the semi-conductor (junction) to be tested. • The measurement current moves from the “+” terminals to the “COM” terminal. It corresponds to the direct testing of the semi-conductor junction. Short-circuit junction: audio indication for a threshold <0.050V Cut or reversed junction (or threshold >3.2V): OL displayed Clamp-on Multimeter Model F03...
  • Page 25: Current Measurements - ( )

    Temperature Measurement - (T°) 4.6.1 Without a Sensor 1. Set the rotary switch to the “T°” position. The temperature displayed is the instrument’s internal tempera- ture (the INT symbol is lit), which is the same as the ambient temperature after a sufficient thermal stabilization time. It can be expressed °C or °F: the unit is chosen with the yellow button. 4.6.2 With a Sensor 1. Connect the sensor (couple K) to the clamp terminals, observ- ing the polarity. 2. Place clamp where the temperature is to be measured. 3. Set the rotary switch to the “T°” position. The temperature dis- played is that of the sensor (EXT symbol is lit). It can be expressed in °C or °F: the unit is chosen with the yellow button. Clamp-on Multimeter Model F03...
  • Page 26: Maintenance

    CHAPTER 5 MAINTENANCE Use only factory specified replacement parts. AEMC will not be held ® responsible for any accident, incident, or malfunction following a repair done other than by its service center or by an approved repair center. Changing the Battery Disconnect the instrument from any source of electricity. 1. Set the switch to OFF. 2. Slide a screwdriver into the slot at the top of the battery cover (rear of the clamp) and push the battery cover upwards.
  • Page 27: Repair And Calibration

    You must contact our Service Center for a Customer Service Authorization Number (CSA#). This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration, or a calibration traceable to N.I.S.T. (Includes calibration certificate plus recorded calibration data). ® Chauvin Arnoux , Inc. ® d.b.a. AEMC Instruments 15 Faraday Drive Dover, NH 03820 USA Tel: (800) 945-2362 (Ext. 360) (603) 749-6434 (Ext. 360) Fax: (603) 742-2346 or (603) 749-6309 repair@aemc.com (Or contact your authorized distributor) Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
  • Page 28: Limited Warranty

    For full and detailed warranty coverage, please read the Warranty Cov- erage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the War- ranty Coverage Information with your records. What AEMC Instruments will do: ®...