4.2.20 Calculation of the Displacement Power Factor (DPF)

Measurement Range (1)
Specified measurement Range
(from 1A
)
AC
Accuracy (2) (3)
Resolution
Note (1) If one of the terms in the calculation of the DPF is displayed as "OL", or
forced to zero, the display of the DPF is an indeterminate value "----".
Note (2) of the previous § apply
Note (3) Measurement stabilization ~8 sec
NOTE: The DPF is always positive.
Specific characteristics in MAX/MIN mode (from 10Hz to 1kHz):

4.2.21 Frequency Measurements

4.2.21.1 Voltage
Measurement
Range (1)
Specified
Measurement
Range
Accuracy
Resolution
4.2.21.2 Current
Measurement Range (1)
Specified Measurement Range
Accuracy
Resolution
Note (1)
In MAX/MIN mode, the operating range is limited to 1kHz.
If the level of the signal is too low (<10% of the range, or V<8V or
I<9A) or if the frequency is less than 5Hz, the device cannot
determine the frequency and displays "----".
Specific Specifications in MAX/MIN mode (from 10Hz to 1kHz):
Uncertainties: add 1% R to the values in the table above.
Capture time: approximately 100ms.
5.0 to
999.9Hz
1 to 100% of the
measurement range
0.1Hz
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.
0.00 to +1.00
0 to 100% of the measurement range
± (5% R +2cts)
1000 to
9999Hz
measurement range
± (0.4% R + 1ct)
1Hz
5.0 to 1999Hz
1 to 100% of the measurement range
± (0.4% R + 1ct)
0.1Hz
75
0.01
10.00 to
19.99kHz
0 to 100% of the
10Hz