AEMC C.A 5205 User Manual

AEMC C.A 5205 User Manual

Digital multimeter

Advertisement

Quick Links

Digital Multimeter

MULTIMETER
Model C.A 5205
ENGLISH
User manual
1

Advertisement

Table of Contents
loading

Summary of Contents for AEMC C.A 5205

  • Page 1: Digital Multimeter

    Digital Multimeter MULTIMETER Model C.A 5205 ENGLISH User manual...
  • Page 2: Safety Precautions

    This device is protected by a double insulation or by a reinforced insulation. No linking is required from the protection earth terminal to ensure the electrical security. Thank you for purchasing a C.A 5205 Multimeter. To get the best service from this instrument : read this user's manual carefully, respect the safety precautions detailed.
  • Page 3: Table Of Contents

    Warranty .................. 7 10 - Maintenance ................7 Technical and Sales Assistance ..........8 1 - PRESENTATION The C.A 5205 is designed for the daily needs of professionals in electricity. - 2000 counts - Automatic recognition of AC and DC...
  • Page 4: Description

    2 - DESCRIPTION LIQUID CRYSTAL DISPLAY 2000 measurement counts (height of digits 0.65") Rate : 2 measurements / second : low battery indication Overload indication : the 1 on the left flashes CONTROL BUTTONS HOLD Hold button freezes the measurement on display RANGE Range selection : Press the range button to change from automatic mode to manual mode.
  • Page 5: Resistance (Ω)

    V DC and AC (1) 200 mV 20 V 200 V 600 V Resolution 0.1 mV 1 mV 10 mV 100 mV Impedance 3 MΩ on DC : ± 1 % reading ± 1 count Accuracy on AC : ± 1 % reading ± 5 counts Permitted 600 V rms and 900 V peak overload...
  • Page 6: General Specifications

    Place the selector switch on the function : - in forward direction the display gives the value of the junction in V (resolution 1 mV) - accuracy : ± 2% rdg ± 15 mV - open circuit voltage : 1.5 V typical - short circuit current : 0.4 mA - in reverse direction the display indicates the open circuit voltage, around 2 V.
  • Page 7: Accessories And Spares (To Order)

    Degree of watertightness (IEC 529) : protection index IP 40 8 - TO ORDER Use the designations and references below. C.A 5205 ............... Cat# 2116.73 Supplied with its shockproof case, a pair of leads with probes attached, two 1.5 V "AA" batteries and this user's manual Accessories and spares AC current probe model MN 251 (200 A AC) ...
  • Page 8: Repair And Calibration

    10-4 Repair and calibration To guarantee that your instrument complies with the factory specifications, we recommend that the Model C.A 5205 be submitted to our factory service center at one-year intervals for recalibration, or as required by other standards.