28
Figure 12: TTL Output Schematic
LVDS Inputs and Outputs (LVDS compliant)
Figure 13: LVDS Input
100Ω
Figure 14: LVDS Output
GPIO Isolation
All of the GPIOs are isolated from the rest of the camera and the camera case. They are
not isolated with respect to each other and have a common return (ground) through pin 5
of the GPIO connector.
Note: The shell connection of the GPIO connector is not isolated and should not be used
as a return (ground) for the GPIO signals. The shell connection is attached to the camera
case.
Programming the GPIO Connector
The connector is programmed through the QuickCam application or through the
QuickCam SDK. After you have installed the QuickCam program, refer to the QuickCam
User's Manual or the QuickCam help topic, GPIO Control, for more information on
programming the connector. Refer to section
information on installing QuickCam.
03-032-10158-09
100Ω
5V
ESD
Protection
Termination: 100 Ω series
Output current: sink 50 mA; source 50 mA
Output voltage: maximum of low 0.55 V @ 32mA; minimum of high 3.8 V @ 32mA.
Spyder 3 GigE User Manual
3 .1 QuickCam Interface for more
2 3 5 H
DALSA