P000 Lock

Notes:
This lock only applies to the handheld programmer: it does not lock access
through communications.
A remote master can change configuration if P799 is set to allow this.
Secures the HydroRanger 200 from changes via the handheld programmer.
Primary Index
Values
Related
WARNING:
Use this lock as backup security only. It uses a fixed value which can
be discovered by unauthorized personnel.
Access this parameter directly (type 000) and enter any value (except 1954) to secure
programming lock. To unlock the HydroRanger 200, access this parameter and enter 1954.

Quick Start (P001 to P007)

P001 Operation
Sets the type of measurement required for the application.
Primary Index
Values
Alters
7ML19985FC03
Global
1954
*
OFF (programming permitted)
Simulation Controls (relays energize based on simu-
-1
lated level)
other
lock activated (programming secured)
Single Point Model
Global
0
Out-of-service
1
Level – how full the vessel is
2
Space – how empty the vessel is
3
*
Distance – distance from transducer to material
4
DPD – dual point difference [6 relay model]
5
DPA – dual point average [6 relay model]
6
OCM – flow rate in an open channel [6 relay model]
7
Pump Totalizer – total pumped volume [6 relay model]
P600 Primary Measuring Device
HydroRanger 200 – INSTRUCTION MANUAL
Simulation on page 83
Dual Point Model
Transducer
Page 117