P275 Scaled Discrete Input Value
Shows the current value of the discrete input after any scaling is applied.
Primary Index
Values
Readings are updated continuously even in PROGRAM mode. The value signals a level
override event.

Standard Data Logging (P300 to P321)

All records can be reset by pressing the CLEAR

Record Temperatures (P300 to P303)

These features display the high and/or low temperatures in °C. When a parameter
relating to a TS-3 Temperature Sensor is accessed, the Point Type display changes to the
TS-3 symbol
.
If the unit is powered up without a temperature sensor connected, the value –50°C is
displayed. This information can help trace problems with both built in and external
temperature sensors.
P300 Temperature, Transducer Maximum
Shows the highest temperature encountered, as measured by the temperature sensor in
the transducer (if applicable).
Primary Index
Values
Related
Press CLEAR
Page 154
Discrete input
Display: view only
Values: dependent on the function of the discrete input
Range of Values
1
0
0 (DI open), 1 (DI closed)
0 (DI closed), 1 (DI open)
Transducer
Range: - 50 to +150°C (view only)
Preset: - 50°C
P301 Temperature, Transducer Min
keys to reset the log after a short circuit on the transducer wiring.
HydroRanger 200 – INSTRUCTION MANUAL
Function (P270)
Forced ON
Forced OFF
Normally Open
Normally Closed
keys.
7ML19985FC03