mA Output Trim (P214 to P215)

This does not affect the P203 value shown, and is used when recalibration of an external
device is impractical.
To adjust the value so that the device correctly indicates 4.00 (when P214 is accessed) or
20.00 mA (when P215 is accessed):
1.
Attach ammeter to HydroRanger 200 mA output.
2.
Access P214, Index 1 (for mA output 1) or 2 (for mA output 2). Press CLEAR and
ENTER
3.
Enter the exact value displayed on the ammeter into P214 (Index 1 or 2).
4.
The ammeter should then read exactly 4.00 mA.
5.
Repeat steps 1 to 4 to set P215, using 20 mA as the desired value.
P214 4 mA Output Trim
Calibrates the 4 mA output.
Primary Index
Values
Related
Adjust this value so the device indicates 4.000 mA when P214 is accessed.
7ML19985FC03
mA output
Range: 0.000 to 22.00
Preset: 3.8
P200 mA Output Range
P213 mA Output Max Limit
mA output
Range: 0.000 to 22.00
Preset: 20.2 mA
P200 mA Output Range / P212 mA Output Min LImit
. The ammeter should show a value near 4 mA.
mA output
Preset: 4.00
P215 20 mA Output Trim