Remote Programming of Output Voltage and Current Limit
Programming Output Current Limit with a 0-10Vdc Source
1. Set jumper JMP2 to its A-B location.
2. Remove the jumpers connecting pins 10 to 11 and 22 to 23 on connector J3.
3. Connect the external current programming source between pins 10 (current limit
Varying the programming voltage from 0 to 10Vdc will cause the output current limit
to vary from 0 to 100% of the model rating. The output voltage is set locally by
adjusting the front panel voltage control.
Note:
Resetting internal jumpers may require a recalibration of the programming circuit to
maintain programming accuracy. See
page

Figure 3.10Programming Output Current Limit with a 0-10Vdc Source

60
WARNING
Exercise caution when resetting internal jumpers. Follow the procedures for
removing the cover and resetting jumpers in
Programming, Sensing, and Monitoring" on page 48
programming input/positive) and 12 (return).
82.
C
B
A
JMP 2
"Configuring for Remote
"Calibrating for Programming Accuracy" on
J3
+
0 - 10 V
Operating Manual for XKW 3kW Series Power Supply
.