Operation
Remote Programming of Output Voltage and Current Limit
Figure 3.4 Programming Output Voltage and Current Limit
Notes:
1. If you are programming both output voltage and current limit with a 0-10V
2. Resetting internal jumpers may require a recalibration of the programming
56
PROGRAMMING SOURCE
OUTPUT
CURRENT LIMIT
PROGRAMMING
SOURCE
SEE NOTE BELOW
WHEN USING 0-10V
(with the REM/LOC Switch)
source and using the REM/LOC switch, set internal jumpers JMP1 and JMP2 to
their A-B locations before connecting the voltage and current limit programming
inputs.
circuit to maintain programming accuracy. See
Accuracy" on page 82
13
12
10
8
25
-
+
OPTIONS
+
0-5V
+
0-1mA
+
0-10V
PROGRAMMING
for the procedures.
Operating Manual for XKW 3kW Series Power Supply
J3
1
14
-
+
PROGRAMMING SOURCE
OPTIONS
+
0-5V
+
0-1mA
+
0-10V
SEE NOTE BELOW
WHEN USING 0-10V
PROGRAMMING
"Calibrating for Programming
OUTPUT
VOLTAGE
PROGRAMMING
SOURCE