I-Class Mark II Maintenance Manual
Advanced Entry Calibration (continued)
Step
Action
Raise the printhead assembly then
proceed according to the media
type:
Die-cut – Remove a label or two
from the liner then position the
liner in the Media Sensor. Adjust
H
the Media Sensor if necessary.
Notched – Position the Media in
the Media Sensor under the notch.
Reflective – Position the Media in
the Media Sensor under the black
mark.
Lower and latch the Printhead
Assembly.
Using the UP and DOWN Arrow
I
buttons set the Gain Number to
00.
Record the reading as a TOF Value
for Gain Number 00 in the table.
Gain Number
Displayed Message
TRAN SENSOR GAIN
TRAN SENSOR GAIN
yyy
Sample Calibration Table
Label Value
00
252
01
250
02
248
...
...
31
09
Adjustments and Maintenance
(1) Do not position the Media
Sensor under a perforation; and
if using preprinted media, ensure
the label area placed over the
31
sensor is free of text, graphics,
(0 - 31)
lines, etc.
(2) Do not move the Media
Sensor after this step.
This is the TOF Value for a gain
setting of 00, where "yyy"
(0 - 31)
represents the current sensor
00
reading.
TOF Value
248
8
Comment
Difference Value