Advanced Entry Calibration - Datamax I-Class Mark II Maintenance Manual

Hide thumbs Also See for I-Class Mark II:
Table of Contents

Advertisement

I-Class Mark II Maintenance Manual
2.1.4

Advanced Entry Calibration

Advanced Entry is an alternate calibration method for special-case media types, where sensor readings
are taken using different sampling algorithms and from a list of these readings the best algorithm is
selected for manual entry into the database.
Advanced Entry Calibration should be used only when Standard Calibration proves unsuccessful.
Calibrate the Media Sensor using the steps below:
The Sensor Calibration menu item is only accessible in the Advanced Menu mode.
Step
Action
A
Turn ON the printer.
Press MENU button. Make sure
MEDIA SETTINGS is highlighted
and press ENTER.
B
Then using the DOWN button,
scroll to SENSOR CALIBRATION
Using the DOWN button, highlight
C
ADVANCED ENTRY then press
ENTER.
Scroll to TRAN SENSOR GAIN (or
REFL SENSOR GAIN, if using
D
reflective media) then press
ENTER.
Install media. Position the Media
in the Media Sensor then lower
E
and latch the Printhead
Assembly.
Adjustments and Maintenance
Displayed Message
READY
(Printer Menu)
ADVANCED ENTRY
TRAN SENSOR GAIN
25
(0 - 31)
TRAN SENSOR GAIN
25
(0 - 31)
6
Comment
Wait briefly for initialization to
complete.
Press EXIT to abort this
procedure.
The following examples detail
die-cut media calibration;
however, unless otherwise
noted, the reflective media
procedure is the same.
See the Operator's Manual for
sensor adjustment instructions.
Do not position the Media
Sensor under a perforation;
and if using preprinted
media, ensure the label
area placed over the sensor
is free of text, graphics,
lines, etc.