13.3 Replacing the RO membrane

Pacific RO 3 – 12:
Pacific RO 20 – 40:
Ansicht von hinten - ohne Rückwand
Back view, with back panel removed
 Unplug the mains plug.
 Remove the back panel.
 Release all hose connections to the pressure tube (1) of the RO-module.
 Take the pressure tube out of the holding sleeves.
 Open the pressure tube cap nut (3) and remove the membrane unit (2).
 Insert the new membrane unit with the two O-rings (4) on the permeate tube in the direc-
tion of the arrow pointing to the pressure tube.
Insertion of RO-membranes the wrong way round would lead to certain
destruction of them!