Table 5-5 Procedure For Configuring Dat Control - Honeywell 9782 Series Operator's Manual

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Step
MENU
1
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV |
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
| PREV |
DISCRETE CONTROL
ON/OFF
PFT
DAT
DIAGNOSTICS
|
DAT CONTROL
2
CYCLE PERIOD
DAT1
DAT2
DIAGNOSTICS
|
ON/OFF CYCLE TIMER 1
3
OUTPUT
INPUT
SETPOINT
PB LIMIT
DIAGNOSTICS
|
7/99
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual

Table 5-5 Procedure for Configuring DAT Control

Screen
Go to CONFIGURATION | DISCRETE CONTROL | DAT.
NEXT
ATTENTION: Any value entered for one of these control-
related parameters will take effect as soon as it is entered.
NEXT
|
Specify the length of the cycle period (in seconds) for the
200 S
time-proportioned control.
Use this screen to select the DAT control output to be
configured.
|
The relay assigned to this DAT control output during I/O
RLY2
setup will already be filled in. (This is read-only
C1
15000 µS
information.)
250 µS
Specify the process value or computed value to be used
|
as the input.
Enter the setpoint at which the output should be turned on.
Enter the PB Limit. This is the value for which the output's
on duration will be maximum.
I/O Setup and System Configuration
Action
5-15