Select valves and fittings with appropriate pressure ratings
Make the process connections as shown in Figure 7-1. Be sure that valves and fittings (S1) have
sufficient pressure ratings to withstand pressure peaks which will occur when process flow is
blocked.
Minimize liquid volume in system
Keep pipe sizes small and couplings close to minimize the liquid volume in the system. Smaller
volumes require less time to rinse.
7/99
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
rinse/cleaning
solution
cell
S1
process
sample
Automatic Wash Setup
Outputs & Alarms Held
Rinse Duration (seconds)
Relay 1 Activated
Resume Delay Time (seconds)
0.588" 0.222"
TIME
Interval Between Rinses
(days or hours)
AutoClean Operation
Figure 7-1 Automatic Cell Wash Setup
9782
to process
S4
to drain
Items outside this area
provided by user
Normal Operation
1.709"
a/n 23185
AutoClean Theory and Piping
7-3