Ce Conformity; Year 2000; Electrical; Power Requirements - Honeywell 9782 Series Operator's Manual

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Conductivity/Resistivity Analyzer/Controller

CE conformity

This product is in conformance with the protection requirements of the following European
Council Directives: 89/336/EEC, the Electromagnetic Compatibility Directive and 73/23/EEC,
the Low Voltage Directive. Conformance of this product with any other "CE Mark" Directive(s)
shall not be assumed.

Year 2000

This instrument will be unaffected by the occurrence of the year 2000.

2.1.4 Electrical

Power Requirements

120 +/- 10% Vac, 47-63 Hz, 15 VA.
240 +/- 10% Vac, 47-63 Hz, 15 VA.
Memory retained by EEPROM when power is off.

Fuse Rating

120V operation : 0.25 amp/250V
240V operation: 0.125 amp/250V

2.1.5 Model Ranges

Standard range models

Conductivity
(microsiemens/cm)
0-1.999*
0-19.99*
0-199.9*
0-1,999
0-19,990
* Equivalent S/m ranges also available.
** For power plant degassed cation conductivity measurements.
*** Recommended cell constant for most power plant applications.
2-2
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Standard Display Ranges: 9782C-S0
TDS**
Resistivity (ohms-cm)
0-1,999 ppb
1-19,990 ppb
0-199.9 ppm
0-1,999 ppm
0-10,000 ppm
Cell Constant Required
0-19.99M
0-1.999M
0-199.9K
0-19.99K
0-1.999k
(cm-1)
0.01
0.1***
1.0
10
25 or 50
7/99