5.9 Selecting a Computed Value to Be Displayed
Purpose
The Analyzer/Controller can calculate and display a computed value. The available values are:
RATIO: ratio between the values measured by the cells;
DIFFERENCE: difference between the values measured by the cells;
%PASSAGE: computed percent passage:
%REJECTION: computed percent rejection:
PPMCO2: computed parts per million carbon dioxide (standard range analyzers only)
Appendix B includes examples of applications using these values.
ATTENTION
If a computed value is used, it is significant which cell is considered "Cell 1" and which is
"Cell 2". This must be taken into account when wiring the inputs. See 6.3 for details.
Procedure
Use the procedure in Table 5-9 to specify a computed value.
Step
MENU
1
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV |
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
| PREV |
ADVANCED FEATURES
CLOCK SETUP
COMPUTED VALUE
AUTO CLEAN
DIAGNOSTICS
| PREV |
7/99
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Table 5-9 Procedure for Specifying a Computed Value
Screen
Go to CONFIGURATION | ADVANCED FEATURES |
COMPUTED VALUE.
NEXT
NEXT
NEXT
I/O Setup and System Configuration
Cell
1
Cell
2
Cell
1
x
100
Cell
2
Cell
1
(
1
)x
100
Cell
2
Action
Cell
1
Cell
2
5-21