Table 5-4 Procedure For Configuring Pft Control - Honeywell 9782 Series Operator's Manual

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Conductivity/Resistivity Analyzer/Controller
Step
MENU
1
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV |
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
| PREV |
DISCRETE CONTROL
ON/OFF
PFT
DAT
DIAGNOSTICS
|
PFT CONTROL
2
DB FREQ RATE
PFT1
PFT2
DIAGNOSTICS
|
PFT 1
3
OUTPUT
INPUT
SETPOINT
PB LIMIT
DIAGNOSTICS
|
5-14
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual

Table 5-4 Procedure for Configuring PFT Control

Screen
Go to CONFIGURATION | DISCRETE CONTROL | PFT.
NEXT
ATTENTION: Any value entered for one of these control-
related parameters will take effect as soon as it is entered.
NEXT
|
Specify the maximum frequency rate (in pulses per
100 P/M
minute) representing 100% output. (Be sure this rate is
compatible with the controlled device.) This rate will be
used for all PFT outputs in the Analyzer/Controller.
|
Use this screen to select the PFT control output to be
configured.
The relay assigned to this PFT control output during I/O
RLY2
setup will already be filled in. (This is read-only
C1
15000 µS
information.)
250 µS
Specify the process value or computed value to be used
|
as the input.
Enter the setpoint at which the output should be turned on.
Enter the PB Limit. This is the value for which the output's
pulse output rate will be at its maximum frequency.
Action
7/99