Conductivity/Resistivity Analyzer/Controller
Table 8-3 Procedure for Performing Calibration Trim Using a Reference Solution
Step
1
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
|
2
CALIBRATION
CELL CAL FACTOR
CAL TRIM 1
CAL TRIM 2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2
DIAGNOSTICS
|
3
CAL TRIM
PUT C1 IN
REFERENCE SOLUTION
ATTENTION
WAIT FOR STABLE
READING!
DIAGNOSTICS
| PREV | NEXT
CAL TRIM C1
4
147.00 µS
DIAGNOSTICS
|
CAL TRIM C1
5
147.00 µS
CAL ADJ FAILED
|
6
8-6
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Screen
MENU
|
|
|
|
Prepare container of the reference solution.
Go to CALIBRATION | CAL TRIM C 1.
Remove the cell from the process.
Rinse the cell thoroughly with de-ionized water.
Immerse it in the reference solution.
Go to the next screen.
The display will show the conductivity of the reference
solution as measured by the cell and Analyzer/Controller
system.
Once the reading is stable, change the value to match the
actual conductivity of the reference solution at its current
temperature.
Press ENTER to save the calibration trim value.
If the calibration trim adjustment is successful, the
calibration menu will again be displayed.
Return the cell to the process and go to Step 6.
If the attempted calibration trim is greater than the
allowable adjustment limit, the operation will not be
successful. The CAL TRIM display will remain on the
screen and an error message will be displayed on the
alarm stripe.
Check the cell and replace if necessary.
After successful calibration, repeat the operation for the
other cell. When calibration trim has been completed for
both cells, return to the main menu.
Action