Table 10-1 Procedure For Calibrating Outputs - Honeywell 9782 Series Operator's Manual

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Conductivity/Resistivity Analyzer/Controller
Step
Screen
1
2
3
4
5
6
7
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV | NEXT
MAINTENANCE
OFFLINE FUNCTIONS
INSTRUMENT SETUP
LEAD RESISTANCE COMP
DIAGNOSTICS
| PREV |
OFF-LINE FUNCTIONS
DISPLAY TEST
KEYBOARD TEST
OUTPUT/RELAY TEST
OUTPUT CALIBRATION
DIAGNOSTICS
|
10-4
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual

Table 10-1 Procedure for Calibrating Outputs

Turn off the power to the Analyzer/Controller. More than
one switch may be required to disconnect power.
With the power off open the case:
Grasp the bottom center portion of the front bezel and pull it
downward and toward you slightly to disengage the bottom of
the bezel from the edge of the case.
Lift the bezel gently to disengage it from the top edge of the
case.
Swing the bezel to the left. (The bezel and display assembly is
mounted on pivot arms.) The safety cover with its warning label
will be visible.
With the power off remove the safety shield by removing the
single screw holding it in place.
Label and remove the field wiring connected to the analog
output terminals.
Feed the test wiring through conduit hole in the case and
connect the appropriate type of meter to the terminals for the
output to be calibrated. Be sure to observe correct polarity.
Close the case and power up the unit.
Go to MAINTENANCE | OFF-LINE FUNCTIONS | OUTPUT
CALIBRATION.
NEXT
|
Action
7/99