Procedure; Table 8-5 Procedure For Entering Tds Conversion Factor - Honeywell 9782 Series Operator's Manual

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Next, perform the calibration. Once calibration has been completed, go back to I/O setup and set
the solution temperature compensation type to the TDS choice, for example "NaCl (TDS)".
At this point you are ready to enter the TDS conversion factor as described in Table 8-5.

Procedure

To enter the TDS conversion factor follow the instructions in in Table 8-5.
Step
1
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
|
CALIBRATION
CELL CAL FACTOR
TDS CVSN FACTOR C1
TDS CVSN FACTOR C2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2
DIAGNOSTICS
|
TDS CONVRSN FACTOR
2
C1
DIAGNOSTICS
| <--
3
TDS CONVRSN FACTOR
ACCEPTED AND SAVED
DIAGNOSTICS
|
4
CALIBRATION
CELL CAL FACTOR
TDS CVSN FACTOR C1
TDS CVSN FACTOR C2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2
DIAGNOSTICS
|
7/99
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual

Table 8-5 Procedure for Entering TDS Conversion Factor

Screen
MENU
|
|
0.5
|
-->
|
|
Go to CALIBRATION | TDS CVSN FACTOR C1.
If you plan to calibrate, do it before entering the
theoretically derived TDS conversion factor (see previous
page).
Enter the appropriate TDS conversion factor for the
selected cell.
If the factor entered is within the permitted range for the
cell constant (see Table 5-10), the Analyzer/Controller will
use the entered value.
If the value is out-of-range, the Analyzer/Controller will
force the value to the appropriate range limit.
Once an acceptable factor has been entered for Cell 1,
repeat the process for Cell 2.
Calibration
Action
8-9