Honeywell 9782 Series Operator's Manual page 124

Two cell conductivity/resistivity analyzer/controller
Table of Contents

Advertisement

Conductivity/Resistivity Analyzer/Controller
Step
PRODUCT INFORMATION
2
SOFTWARE VER:
MODEL NUMBER:
9782X-nn-Xn-XXXX
DIAGNOSTICS
EDIT |
3
FACTORY PASSWORD
ENTER PASSWORD
DIAGNOSTICS
|
MODEL NUMBER:
4
9782X-nn-Xn-XXXX
DIAGNOSTICS
|
1
If you do not have the factory password, call the Technical Assistance Center at 1-800-423-9883.
10-8
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator's Manual
Screen
The software version is read-only information.
A1.0
If the model number displayed no longer matches the
hardware because you have added or changed an option
card, use the specified function key to edit the model
|
number. (See Section 2 for available model number
selections.)
Note that a special password is required to change the
model number. (This is not the same as the security
OOOO
password described in 10.7.) Enter it when prompted.
This password is included with the instructions for installing
|
the new card.
ATTENTION: Changing the model number stored in
memory will do a "cold reset" of the unit, returning ALL
configuration values to the factory settings shown in Table
5-1. During a cold reset some I/O setup values will also
be returned to the defaults, and the cal factor will be reset
|
to match the cell constant. The calibration trim or TDS
conversion factor will be removed.
The "X" values in the model number are "don't care"
conditions, because these characters do not affect I/O
setup and configuration parameters.
Once you have completed the model number change,
press ENTER to accept the change.
To cancel the edit and avoid resetting the unit, press the
MENU key.
Action
1
7/99