(8) Clicking the OK button begins calibration.
(9) Calibration ends after approximately 30 minutes.
Click the OK button to complete calibration.
(10) To check the calibration results, click the Confirm button in the Calibration menu (shown
in step (5) above).
(11) The following message is displayed. Insert the calibration sheet into the scanner as
instructed.
(12) Click the OK button to start scanning.
The scanned data is displayed when scanning is complete. Colors may differ slightly
for individual sensors to make it easier to identify problem areas in calibration. This
does not indicate a defect.
Check that there are no vertical streaks, such as white patches, in the scanned data.
(Streaks occur when calibration is not performed correctly due to contamination by dust
or dirt.)
(13) If the data is normal, calibration is complete.
Click the Close button and exit the Scanner Adjustment Program.
CS2000
5. DAILY MAINTENANCE
5 – 11