Installing A Network Card In An Agilent Module - Agilent Technologies 1100 Series User Manual

Lc devices connecting the hardware and triggering data acquisition
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Connecting the Hardware

Installing a Network Card in an Agilent Module

Installing a Network Card in an Agilent Module
4
Agilent 1100 Series LC Devices Connecting the Hardware and Triggering Data Acquisition
For the Agilent LC system to communicate with the Xcalibur data system through an
Ethernet connection, one of the modules in the Agilent 1100 Series LC stack must have a
network interface card. If the stack has a detector, the card is usually installed in the detector.
Tip Vendors label network cards with a unique MAC address, which is a 12-digit
hexadecimal number. The first portion of the MAC address identifies the vendor. The
second portion of the MAC address is a unique serial number. For example, a typical mark
on a network card is AD: 0030C1C49982, in which 0030C1C49982 is the stack MAC
address. The first portion, 0030C1, identifies the vendor as Hewlett-Packard. The MAC
address of the HP JetDirect 400N network card is located on the back of the card.
Because you need the MAC address when you configure the Agilent stack, record the
network card MAC address before you install the card.
Agilent 1100 Series LC systems with JetDirect® network cards must have revision 2
mainboards to function properly with LAN communications.
revision level for each module. The serial number listed in the table and all serial numbers
after that number are supported.
Table 3. Supported serial number
Module
G1310A
Isocratic pump
G1311A
Quaternary pump
G1312A
Binary pump
G1313A
ALS autosampler
G1314A
Variable wavelength detector
G1315A
Diode array detector
*
All serial numbers above the listed number in numeric order are supported
P/N (Mainboard)
Serial Numbers for Supported Version
G1311-65520
DE64300355 (made in Germany)
US64400233 (made in U.S.A.)
G1311-65520
DE64301137 (made in Germany)
US64401134 (made in U.S.A.)
G1312-65520
DE64300703 (made in Germany)
US64400425 (made in U.S.A.)
G1313-65520
DE64302092 (made in Germany)
US64400886 (made in U.S.A.)
G1314-65520
JP64201926 (made in Japan)
G131-65520
DE64301532 (made in Germany
US64400333 (made in U.S.A.)
Table 3
shows the required
Thermo Scientific
*

Advertisement

Table of Contents
loading