4.3.7 GC Calibration
Connect a 4 terminal load resistor (0.1 ohms) between the HIGH/LOW DRIVER and
HIGH/LOW SENSE of the Guardian 5000. Connect the 30 A AC ammeter in series with the
HIGH DRIVER terminal.
Press
Display
Press
Display
Press
Press
Press
Press
Display
Press
Display
Press
Press
Press
Change the load resistor to 50 mΩ and connect the AC voltmeter between HIGH and LOW
SENSE.
Press
Display
Press
Display
Press
Press
Press
Change the load resistor to 100 mΩ
Press
Display
Press
Display
Press
Press
Press
www.valuetronics.com
[UP] or [DOWN] key as necessary to display
CL
3.0
OFST
[OFF] key to display
1
3.0A 200.0 0.0 sec
[STOP] [START]
[2] [.] [8] [7] [ENTER]
[STOP]
[DOWN] key to display
CL
25.0
FULL GrA
[OFF] key to display
1
25.0A 200.0 0.0 sec
[STOP] [START]
[2] [4] [.] [8] [7] [ENTER]
[STOP]
[DOWN] key to display
CL
5.0
OFST
[OFF] key to display
1
5.0A 200.0 0.0 sec
[STOP] [START]
[0] [.] [2] [5] [7] [ENTER]
[STOP]
[DOWN] key to display
CL
30.0
FULL GrV ; 30A into 100mohm
[OFF] key to display
1
30.0A 200.0 0.0 sec
[STOP] [START]
[2] [.] [8] [4] [7] [ENTER]
[STOP]
GrA
; read out the ammeter value
; example 2.87A
; stop GrA offset calibration
; read out the ammeter value
; example 24.87A
; stop GrA full scale calibration
GrV ; 5.0A into 50mohm
; read out the voltmeter value
; example 0.257 volts
; stop GrA offset calibration
; read out the voltmeter value
; example 2.847 volts
; stop GrA full scale calibration