9. HILI (?) {f | *}
Set the high limit value according to mode selected. * disables the high limit.
In Hipot mode if the High Limit/2 is smaller than the Low Limit, the low limit is
disabled.
In IR mode the value can not be smaller than the Low Limit, otherwise error 2 is
indicated.
In Ground Continuity mode the maximum high resistance limit is 510mΩ or
R = 6.3 V/I (i.e. 210mΩ for 30 Amps).
10. LOLI (?) {f | *}
Set the low limit value according to mode selected. * disables the low limit.
In Hipot mode the limit is disabled if greater than High Limit/2.
In IR mode the limit is disabled if greater than the High Limit.
In Ground Continuity mode the limit can not be set.
11. SARC (?) {f | *}
Set the arc limit value (Hipot mode only). * disables the limit.
12. HICH (?) {n | *}
Set the High Channel from 1 - 8. * disables the limit. More than one channel is
possible (Ex: HICH 1 3 5).
13. LOCH (?) {n | *}
Set the Low Channel from 1 - 8. * disables the limit. More than one channel is
possible (Ex: LOCH 1 3 5). Can not be set in Ground Continuity mode.
14. TIME (?) {f | *}
Set the Test Time from 0.1 - 99.9 sec. * disables the time.
15. RAMP (?) {f | *}
Set the voltage rise time from 0 - 99.9 sec for Hipot mode. * disables the time.
www.valuetronics.com
Ground Continuity
f = 0.1 - 500.0 mΩ
AC Hipot
f = 0.01 - 40.00 mA
DC Hipot
f = 0.01 - 20.00 mA
Insulation Resistance f = 1 - 9999 MΩ
Ground Continuity
f = 0.1 - 500.0 mΩ
AC Hipot
f = 0.01 - 40.00 mA
DC Hipot
f = 0.01 - 20.00 mA
Insulation Resistance f = 1 - 9999 MΩ
AC Hipot
f = 0.01 - 40.00 mA
DC Hipot
f = 0.01 - 20.00 mA
Maintenance & Calibration 3-7