4.3.6 DC Current Calibration

Connect a load resistor (10 Mohms) between the Hipot H.V. OUTPUT of the Guardian 5000
and the High terminal of the DC ammeter. Connect the Low terminal of the DC ammeter to
the LOW DRIVER terminal (black) of the Guardian 5000.
Press
Display
Press
Display
Press
Press
Press
Press
Display
Press
Display
Press
Press
Press
Press
Display
Press
Display
Press
Press
Press
Press
Display
Press
Display
Press
Press
Press
Maintenance & Calibration 4-6
www.valuetronics.com
[UP] or [DOWN] key as necessary to display
CL
OFST
2.999
[OFF] key to display
1
1.2KV
2.999mA
[STOP] [START]
[0] [.] [1] [2] [4] [ENTER]
[STOP]
[DOWN] key to display
CL
FULL 2.999
[OFF] key to display
1
1.2KV
2.999mA
[STOP] [START]
[2] [.] [9] [0] [3] [ENTER]
[STOP]
[DOWN] key to display
CL
OFST
15.0
[OFF] key to display
1
1.2KV
15.00mA
[STOP] [START]
[2] [.] [9] [0] [3] [ENTER]
[STOP]
[DOWN] key to display
CL
FULL 15.0
[OFF] key to display
1
1.2KV
15.00mA
[STOP] [START]
[1] [4] [.] [0] [2] [ENTER]
[STOP]
DCA
0.0 sec
; read out the ammeter value
; example 0.124mA
; stop DCA 2.99mA range offset
calibration and change the dummy load
resistor to 420 kohms
DCA
0.0 sec
; read out the ammeter value*
; example 2.903A
; stop DCA 2.99mA range full scale
calibration
DCA
0.0 sec
; read out the ammeter value
; example 2.903mA
; stop DCA 15.0mA range offset
calibration and change the dummy load
resistor to 80 kohms
DCA
0.0 sec
; read out the ammeter value*
; example 14.02mA
; stop DCA 15mA range full scale
calibration