4.3.3 DC Voltage Calibration
Connect the DC voltmeter between the H.V. OUTPUT and the LOW DRIVER terminal.
Press
Display
Press
Display
Press
Press
Press
Press
Display
Press
Display
Press
Press
Press
4.3.4 IR Voltage Calibration
Connect the DC voltmeter between the H.V. OUTPUT and the LOW DRIVER terminal.
Press
Display
Press
Display
Press
Press
Press
Press
Display
Press
Display
Press
Press
Press
Maintenance & Calibration 4-4
www.valuetronics.com
[UP] or [DOWN] key as necessary to display
CL
50
OFST
[OFF] key to display
1
0.05KV
.500mA
[STOP] [START]
[0] [.] [0] [6] [2] [ENTER]
[STOP]
[DOWN] key to display
CL
4000 FULL DCV
[OFF] key to display
1
4.00KV
.500mA
[STOP] [START]
[4] [.] [0] [5] [2] [ENTER]
[STOP]
[UP] or [DOWN] key as necessary to display
CL
50
OFST
[OFF] key to display
1
50V
1MΩ
[STOP] [START]
[6] [2] [ENTER]
[STOP]
[DOWN] key to display
CL
1000 FULL IR
[OFF] key to display
1
1000V 1MΩ
[STOP] [START]
[1] [0] [5] [2] [ENTER]
[STOP]
DCV
0.0 sec
; read out the HV meter value
; example 0.062KV
; stop DCV offset calibration
0.0 sec
; read out the HV meter value
; example 4.052KV
; stop DCV full scale calibration
IR
0.0 sec
; read out the HV meter value
; example 62V
; stop IR voltage offset calibration
0.0 sec
; read out the HV meter value
; example 1052V
; stop IR full scale calibration