1
INTRODUCTION
User's Manual
The Planmeca ProMax 3D Max X-ray unit uses cone
beam
volumetric
dimensional X-ray images. The images can be used for
examination of dentomaxillofacial, ear, nose and throat,
as well as other cranial anatomies. The X-ray unit is
allowed to be used only under supervision of a dental /
health care professional.
This manual describes how to operate the Planmeca
ProMax 3D Max X-ray unit. Please read these instructions
thoroughly before using the unit.
You need a PC with the Planmeca Romexis imaging
software program in order to save, view and modify the
3D images. The Planmeca Romexis software program
has a separate manual that should be used in conjunction
with this manual.
The Planmeca ProMax 3D Max X-ray unit fulfills the
requirements of Directive 93 / 42 / EEC (Class IIb).
0537
The Planmeca ProMax 3D Max X-ray unit fulfills the
requirements of standard EN 55011, class A.
CAUTION FOR US USERS:
Federal law restricts this device to sale by or
on the order of a dentist.
NOTE The X-ray unit's software revision is shown briefly on
the control panel when the unit is switched on. This
manual is valid for software revision 2.2.0.0.r or later.
This software revision is compatible with Romexis
software revision 2.7.0.r or later.
The display values shown in this manual are only
examples and should not be interpreted as recommended
values unless otherwise stated.
The Planmeca ProMax 3D Max X-ray unit features preset
exposure values. The exposure values have been preset
at the factory to automatically correspond with the
selected patient size and image resolution. However, the
exposure values required to produce good X-ray images
will vary considerably according to the build and age of
the patient. The preset exposure values are average
values and they are only meant to guide the user. Users
are encouraged to develop their own radiographic
techniques based on these values. The kV value can be
reduced from the suggested value to improve image
contrast. The radiation dose can be reduced by
decreasing the suggested mA value.
Make sure that you are fully acquainted with the
appropriate radiation protection measures and these
instructions before using the X-ray unit.