Test exposure (i460)
Network settings (i480)
User's Manual
Select Test exposure (i460) on the Special functions
(i400) display. The display shown below appears.
This option allows you to take a test exposure. The Test
exposure (i460) display shows the exposure values which
are in use on the main display.
Touch the arrows to select the exposure values (kV, mA
and time) you wish to use. Then touch the Test field. The
word Ready will appear at the bottom of the display and
you can now press the exposure button to take a test
exposure. Note that the C-arm will not move when you
take a test exposure.
NOTE If the X-ray unit has not been used for a week or more
and you receive error message E332 (Arcing across X-
ray tube), take four test exposures with a one-minute
interval between exposures. Use the lowest possible
kV and mA values and a long exposure time. If the
error message reappears, contact your service
technician for help.
Refer to the Technical Manual for further information on
network settings.