Input Probe System - HP 3563A Operating Manual

Control systems analyzer
Table of Contents

Advertisement

In put Probe System
The digital probe system consists of probes, pods, probe cables, and grabbers. The input probe leads
contain an RC network consisting of a
cables are shielded resistive transmission line. Each cable, probe, and pod assembly is (collectively)
4.5 feet long and contains 16 data channels, one clock channel, and a pod ground. See the
Installation Guide for labeling instructions.
Caution
The probe grounds are chassis (earth) grounds, not "floating" grounds.
I
Each input probe is a 12-inch twisted-pair cable connected to the cable at the pod. One end of the
probe contains an input RC network and a lead to connect to the target system. The other end of
the probe has a two-pin connector that plugs into the pod.
You can connect the probe directly to test pins on your target system or to grabbers. Grabbers are
spring-loaded connecting hooks designed to fit on adjacent IC pins. See figure 13-5. If you want to
connect to test pins without using the grabbers, the pins must be 0.63 mm (0.025 in.) square or round
pins with a diameter between 0.66 mm (0.026 in.) and 0.84 mm (0.33 in.). Most IC test clips work.
90.9
kQ resistor in parallel with an 8 pF capacitor. The input
Digital (.;onnectlons
t
Digi
al Overview: The Hardware
1 3-5

Advertisement