Using The Time Capture Mode
Using Capture Blocks As Throughput Files
U s i n g Capture Blocks As Throug hput Files
Captured data blocks can be used as input to a throughput measurement, if they are saved to disc.
To do this, first display the capture block using TIME BUFFER in the VIEW INPUT menu. Next,
save this display to disc using SAVE FILE in the DISC menu. "Using Capture Blocks as Throughput
Files," the corresponding section in chapter 9, explains how to identify the saved capture data as the
active file for throughput. For more information on using throughput, refer to chapter 9. For more
information on saving and recalling disc files, see chapter 1 1.
Note
If you capture digital data with the HP 3563A, then store it to disc and analyze it
with the HP 3562A, the HP 3562A scales the amplitude according to the following
ratio:
(32768126028.55)
Scaling is done when the HP 3562A reads the capture file. Thus, the data in the
capture file is not altered.
6-28
=
1 .258925