Using The Time Capture Mode

Measurement Setup: Time Capture Mode

M easurement Setup: Ti m e Capture Mode
The steps necessary to set up a measurement in time-capture mode are different than those used for
linear resolution or log resolution and swept-sine modes. To perform a time-capture measurement,
you first capture the data (to the analyzer's internal memory), then measure the captured data. The
remainder of this chapter details the steps necessary to set up a time capture measurement, which
are as follows:
Capture the Data
1.
Connect the analyzer to your device under test. See chapter
connections; see chapter
measurement.
Preset the analyzer (optional): press
2.
options).
3.
Select time-capture mode: press
4.
Configure the input: press
5.
View the input signal (if necessary): press
6.
Set up the source (if necessary): press
SPCL FCTN
menu.
7. Set the input range: press
8.
Select the capture channel: press
9. Set the capture length: press
Set the time length or frequency span: press
10.
1 1 .
Set up triggering: press
Start the capture: press
12.
13.
View the captured data: press
TIME RECORD
and
14.
View the capture header (if desired): press
CAPTUR HEADER.
6-2
14
for details on planning a digital or mixed (analog and digital)
PRESET
M EAS MODE
I NPUT CONFIG.
VIEW I NPUT .
SOURCE.
RANGE.
(See chapter 7, "Input Setup," for details).
SELECT M EAS
M EAS MODE, CAPTUR SELECT, CAPTUR LENGTH.
SELECT TRIG.
(See chapter 7, "Input Setup," for details).
MEAS MODE, CAPTUR SELECT, START CAPTUR.
VIEW INPUT
LINEAR S PEC).
M EAS MODE, CAPTUR SELECT
13
for details on digital
(see chapter
for a description of the preset
12
followed by
TIME CAPTUR.
(See chapter 7 for details).
If necessary, set up source protection in the
followed by
C H 1 ACTIVE
FREQ.
(use the capture pointer and
CH 2 ACTIVE.
or
TIME BUFFER,
followed by