System Leds - Renishaw TONiC T20x1 REXM20 Installation Manual

Ultra-high accuracy angle encoder system
Table of Contents

Advertisement

System LEDs

T20x1 readhead LED diagnostics
LED
Indication
Green
Orange
Incremental
Red
Set-up
Green (flash)
Reference
Orange (flash)
mark
Red (flash)
On
Operating
Off
Single-flashing
CAL
Calibration
Double-flashing
Flashing at
Reset
power-up (< 2s)
1
The flash will effectively be invisible when passing the reference mark if the incremental signal is > 70%.
2
The alarm output will take the form of 3-state or line driven E signal depending on DSi configuration. The Ti interfaces connected to the DSi must be configured with the line driven alarm format.
The alarm output is a momentary status only. It is only present while the fault condition remains.
The alarm may result in axis position error, redatum to continue.
3
Some Ti interfaces do not output overspeed alarm. See the TONiC encoder system data sheet (Renishaw part no. L-9517-9337) for details of interface configuration. This can be downloaded from our website at
www.renishaw.com/tonicdownloads
and is also available from your local Renishaw representative.
34
TONiC
REXM20 / REXT20 angle encoder system installation guide
Status
Normal set-up: signal level > 70%
Acceptable set-up; signal level
50% to 70%
Poor set-up; signal may be too
low for reliable operation; signal
level < 50%
1
Normal phasing
Acceptable phasing
Poor phasing; clean the ring and
recalibrate if required
Automatic Gain Control – On
Automatic Gain Control – Off
Calibrating incremental signals
Calibrating reference mark
Restore factory defaults
Ti interface LED diagnostics
Signal
Indication
Normal setup; signal level 110%
Purple
to 135%
Optimum setup; signal level 90%
Blue
to 110%
Normal set-up: signal level 70%
Green
to 90%
Acceptable set-up; signal level
Orange
50% to 70%
Incremental
Poor set-up; signal may be too
Red
low for reliable operation; signal
level < 50%
Red / blank -
Poor set-up; signal level < 20%;
flashing
system in error
Blue / blank -
Overspeed
flashing
Purple / blank -
Over signal; system in error
flashing
Reference
Reference mark detected (speed
Blank flash
mark
< 100 mm/s only)
Alarm
Status
output
2
No
No
No
No
No
Yes
3
; system in error
Yes
Yes
No

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tonic t20x1 rext20

Table of Contents