Renishaw Tonic Installation Manual

Renishaw Tonic Installation Manual

Dsi dual readhead rotary encoder system

Advertisement

Quick Links

Installation guide
M-9653-9298-01-A
RSLM high accuracy linear encoder
DSi dual readhead rotary encoder system

Advertisement

Table of Contents
loading

Summary of Contents for Renishaw Tonic

  • Page 1 Installation guide M-9653-9298-01-A RSLM high accuracy linear encoder DSi dual readhead rotary encoder system...
  • Page 2 Renishaw plc. The user is cautioned that any changes or modifi cations not expressly approved by Renishaw plc or authorised representative could void the user’s authority to operate the equipment.
  • Page 3: Storage And Handling

    Storage and handling System Readhead only N-heptane Acetone Chlorinated COCH Solvents Propan-2-ol Methylated CHOHCH Spirits Storage Operating Humidity System +70 °C 0 °C System +70 °C Rated up to +40 °C, -20 °C 95% relative humidity (non-condensing)
  • Page 4 DSi installation drawing Dimensions and tolerances in mm 17.4 17.4 26.5 26.5 42.5 17.5 Ø3.5 through 2 positions 21.4 LED location Access to propoZ button 92.5 19.4...
  • Page 5: Summary Of Procedures

    When you first install the DSi: Refer to the REXM/T Installation guide Select the correct head orientation switch position (M-9653-9248) for details. Install and calibrate each TONiC This switch can be found on the rear of When you first install the DSi: encoder separately.
  • Page 6 Calibration The system must be calibrated if the installation is disturbed or if the DSi has been moved to a different axis. 1) Install the TONiC readheads as detailed in the TONiC REXM encoder system Remote mounting installation guide (M-9653-9428).
  • Page 7 Selecting a new propoZ position The propoZ position can be re-programmed as many times as the user requires. Ensure the DSi is calibrated and has completed the initialisation routine. You will notice that the LED on the DSi flashes ‘blank’ as it passes the last programmed propoZ position. To change the propoZ position…...
  • Page 8 Open collector routine when the DSi was fi rst installed. This could be caused by a Initialization status miscount from one of the TONiC encoders. If the pulse count alarm is – Inner Not connected...
  • Page 9: General Specifications

    TONiC interface for remote versions. 5 VS 0 VS The TONiC and DSi encoder system conforms to the relevant harmonised European standards for electromagnetic compatibility, but must be correctly integrated to achieve EMC compliance. External In particular, attention to shielding arrangements is essential.
  • Page 10 *M-9653-9298-01* RENISHAW ® and the probe emblem used in the RENISHAW logo are registered trademarks of Renishaw plc in the UK and other countries. apply innovation is a trademark of Renishaw plc. 2010-2011 Renishaw plc All rights reserved Issued 0711...