Renishaw TONiC T20x1 REXM20 Installation Manual page 39

Ultra-high accuracy angle encoder system
Table of Contents

Advertisement

TONiC system and remote DSi
Inner shield
TONiC readhead
TONiC readhead
Outer shield
IMPORTANT: The outer shield should be connected to the machine earth (Field Ground). The inner shield should be connected to 0 V at the receiving electronics only. Care should be taken to ensure that
the inner and outer shields are insulated from each other. If the inner and outer shields are connected together, this will cause a short between 0 V and earth, which could cause electrical noise issues
NOTE: Maximum cable lengths as detailed in
Ti interface
Extension cable
Ti interface
Extension cable
'General specifications'
on page 43.
DSi
Extension cable
Customer
electronics
5 V
Output signals
0 V
www.renishaw.com
39