System connection

Approved ESD precautions must be followed at all times during readhead
and interface electrical connections.
The readhead is connected to the Ti interface via a small, rugged
connector to allow for easy feed-through during installation.
Connecting the readhead
u
Remove the cover plate as shown (2 x M2.5 hex head screws).
u
Taking care not to touch the pins, plug the connector into the socket in
the interface, ensuring correct orientation as shown.
u
Refit the cover plate ensuring the cable ferrule is located in the recess
on the inside and no wires are trapped under the cover plate.
NOTE: The tightening torque should be between 0.25 Nm and 0.4 Nm.
Disconnecting the readhead
u
Remove the cover plate on the interface
(2 x M2.5 hex head screws).
u
Gently lever the connector PCB
(on the end of the cable) out of the socket.
u
Place the connector in an anti-static bag.
u
Refit the cover plate.