12.1 CABLIBRATION S/W INSTALLATION
Please read the following calibration instruction before
starting calibration software with monitors.
▪ Avoid shocks or vibrations when calibrating TVLogic monitors.
These may damage the unit and cause it to malfunction.
▪ Wait minimum 20 minutes before starting calibration with
TV Logic monitor
▪ Start calibration when the brightness change of the screen is less
than 1cd/m².
▪ Recommendable Room temperature for calibrating monitor is
20°C ± 5°C
▪ Recommendable Room (Environmental) brightness is less than
5 cd/m². While calibrating Monitor, light from outside should be
cut off
▪ Do not expose this monitor directly to the sun while calibrating
TVLogic monitors