Propaq M Overview
This chapter includes step-by-step procedures for various maintenance tests. At a minimum, a
trained biomedical technician must perform a CO
once per year. A CO
calibration. It is necessary to perform maintenance tests after a module is repaired to ensure that
the functions of the Propaq M unit work properly. Use the checklist at the back of this
document (ZOLL Propaq M Maintenance Tests Checklist) to record your results of the
maintenance tests.
Because the Propaq M units must be maintained ready for immediate use, it is important for
users to conduct the Operator's Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL Propaq M Operator's Guide for the Operator's Shift Checklist.)
9650-000821-01 Rev. C
Maintenance Tests
calibration check may be performed at any time to verify the CO
2
Propaq M Service Manual
Chapter 1
calibration and an NIBP calibration check
2
2
5