Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

12 Advantage Windows

A power cord is supplied with Advantage Windows. It is connected to outlet fed by the main
disconnect room device.
Verify that the ground between AW, AW spliter and monitors ground are equipotential
(0.1 ohm equipotentiallity between room ground point (busbar) and wall outlets).
To power the AW splitter, use the appropiate cable out of lot P/N 2385173.
178
Direction 5314528-2-1EN, Revision 1
12 Advantage Windows